A Validated Enantioselective HPLC Method for Assay of S-Amlodipine Using Crown Ether as a Chiral Stationary Phase

Research Authors
Ahmed M. Abdel-Megied, Samia M. El-Gizawy, Osama H. Abdelmageed, Mahmoud A. Omar, Sayed M. Derayea, Hassan Y. Aboul-Enein
Research Journal
Current Analytical Chemistry, DOI:10.2174/1573411012666160527151803
Research Publisher
NULL
Research Rank
1
Research Vol
Vol. 13, No. 2
Research Website
NULL
Research Year
2017
Research Member
Research Abstract

Background: Hypertension represents a widespread cardiovascular condition although usually asymptomatic, is a major risk factor for myocardial ischemia, renal failure and stroke. Amlodipine is considered a powerful dihydropyridine calcium channel blocker used as antihypertensive agent. Its mechanism of action depend on relaxing the smooth muscle of the artery wall and decrease peripheral resistance.
Methods: An accurate, sensitive and robust chiral HPLC method was developed for separation and quantitation of S(-)- amlodipine simultaneously in the presence of its R(+)- isomer. Chiral separation was applied using Daicel CROWNPAK CR(+) (5 µm, 4.0x150 mm) column which contains (3,3'-diphenyl-1,1'-binaphthyl)-20-crown-6-ether coated into a 5 µm reversed phase silica support. The mobile phase system was aqueous acidic 70% HClO4 (pH 2.0) and methanol in the proportion of (95:5 v/v), filtered through 0.45 µm membrane and degassed by before use, pumped at a flow rate was 0.8 mL min-1 with UV detector adjusted at 238 nm.
Results: The chromatographic HPLC method was validated with respect to ICH guidelines. Linear concentration range was 5-60 µg mL-1 with correlation coefficient (r) about 0.9998, the detection and quantitation limits was found 1.66 and 5.05 µg m-1; respectively. Accuracy was evaluated using standard addition technique with mean recovery about 98.40% while precision was assessed at intraday and interday level. It was found that all %RSD values below 2%
Conclusion: A simple, enantioselective, chiral HPLC method was developed and validated for the quantitation of amlodipine [S-(+)- isomer] in racemic tablets using CROWNPAK CR(+) chiral stationary phase. The proposed method is specific, precise, accurate, and robust which can be successfully applied for the routine analysis of S-AML in bulk and pharmaceutical dosage formulations.