Measurements and Calculation for Different Electron beam Shielding Block Shapes in Therapeutic LINAC

مؤلف البحث
- Zienab E Mounir, Galal S. Hassan, Ahmed S. Ahmed, and Mostafa A. Hashem
المشارك في البحث
سنة البحث
2003
مجلة البحث
Arab Journal of Nuclear Sciences and Applications
الناشر
NULL
عدد البحث
36
تصنيف البحث
4
صفحات البحث
319-324
موقع البحث
NULL
ملخص البحث

The filed shape must be frequently treated in terms of shielding block choices in man clinical applications of electron beam therapy. Because of the difficult of measuring the output factors for individual rectangular electron beam field for each patient, it is necessary to know the output factors for rectangular field size electron beam fields to accurately deliver the prescribed dose to the target. The output factors for rectangular fields have been found that the equivalent area technique generally used in X-ray treatment can be utilized reasonably well to predict the output factors for rectangular electron fiel