A sensitive and selective voltammetric method of analysis is developed for determination of trace amounts of Cr(V1) ions in neutral nitrate phosphate media This method is based on controlled adsorptive preconcentration of chromium species at Hanging
Mercury Drop Electrode (HMDE). The adsorptive stripping response was evaluated with respect to preconcentration time and potential and dependence on composition of
supporting electrolyte As low as SxIO-’ mol L” (0 25 pg/L) and lx109 mol L’ (0.52 yg/L) Cr(V1) can be detected using Cathodic Linear Sweep Stripping Voltammetry (CLSSV) and Differential Pulse Cathodic Stripping Voltammetry (DPCSV), respectively.
Research Abstract
Research Department
Research Journal
J. Analytical Letters,
Research Member
Research Publisher
NULL
Research Rank
1
Research Vol
29(8) (1996) 1431- 1445.
Research Website
NULL
Research Year
1996
Research Pages
15